![Agilent Technologies InfiniiVision 3000 3024A MSO-X User Manual Download Page 145](http://html1.mh-extra.com/html/agilent-technologies/infiniivision-3000-3024a-mso-x/infiniivision-3000-3024a-mso-x_user-manual_2868679145.webp)
Triggers
10
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
145
Using AutoScale
to Set Up Edge
Triggers
The easiest way to set up an Edge trigger on a waveform is to use
AutoScale. Simply press the
[AutoScale]
key and the oscilloscope will
attempt to trigger on the waveform using a simple Edge trigger type. See
"Use Auto Scale"
on page 30.
N O T E
Alternating edge mode is useful when you want to trigger on both edges of a clock (for
example, DDR signals).
Either edge mode is useful when you want to trigger on any activity of a selected source.
All modes operate up to the bandwidth of the oscilloscope except Either edge mode, which
has a limitation. Either edge mode will trigger on Constant Wave signals up to 100 MHz, but
can trigger on isolated pulses down to 1/(2*oscilloscope's bandwidth).
N O T E
MegaZoom Technology Simplifies Triggering
With the built-in MegaZoom technology, you can simply AutoScale the waveforms, then
stop the oscilloscope to capture a waveform. You can then pan and zoom through the data
using the Horizontal and Vertical knobs to find a stable trigger point. AutoScale often
produces a triggered display.
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...