![Agilent Technologies InfiniiVision 3000 3024A MSO-X User Manual Download Page 63](http://html1.mh-extra.com/html/agilent-technologies/infiniivision-3000-3024a-mso-x/infiniivision-3000-3024a-mso-x_user-manual_2868679063.webp)
Vertical Controls
3
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
63
To adjust the vertical scale
1
Turn the large knob above the channel key marked
to set the
vertical scale (volts/division) for the channel.
The vertical scale knob changes the analog channel scale in a 1- 2- 5 step
sequence (with a 1:1 probe attached) unless fine adjustment is enabled
(see
"To change the vertical scale knob's coarse/fine adjustment setting"
on
page 65).
The analog channel Volts/Div value is displayed in the status line.
The default mode for expanding the signal when you turn the
volts/division knob is vertical expansion about the ground level of the
channel; however, you can change this to expand about the center of the
display. See
"To choose "expand about" center or ground"
on page 289.
To adjust the vertical position
1
Turn the small vertical position knob ( ) to move the channel's
waveform up or down on the display.
The voltage value momentarily displayed in the upper right portion of the
display represents the voltage difference between the vertical center of the
display and the ground level (
) icon. It also represents the voltage at
the vertical center of the display if vertical expansion is set to expand
about ground (see
"To choose "expand about" center or ground"
on
page 289).
To specify channel coupling
Coupling changes the channel's input coupling to either
AC
(alternating
current) or
DC
(direct current).
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...