![Agilent Technologies InfiniiVision 3000 3024A MSO-X User Manual Download Page 229](http://html1.mh-extra.com/html/agilent-technologies/infiniivision-3000-3024a-mso-x/infiniivision-3000-3024a-mso-x_user-manual_2868679229.webp)
Measurements
14
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
229
Rise Time
The rise time of a signal is the time difference between the crossing of the
lower threshold and the crossing of the upper threshold for a
positive- going edge. The X cursor shows the edge being measured. For
maximum measurement accuracy, set the horizontal time/div as fast as
possible while leaving the complete rising edge of the waveform on the
display. The Y cursors show the lower and upper threshold points.
Fall Time
The fall time of a signal is the time difference between the crossing of the
upper threshold and the crossing of the lower threshold for a
negative- going edge. The X cursor shows the edge being measured. For
maximum measurement accuracy, set the horizontal time/div as fast as
possible while leaving the complete falling edge of the waveform on the
display. The Y cursors show the lower and upper threshold points.
Delay
Delay measures the time difference from the selected edge on source 1
and the selected edge on source 2 closest to the trigger reference point at
the middle threshold points on the waveforms. Negative delay values
indicate that the selected edge of source 1 occurred after the selected
edge of source 2.
1
Press the
[Meas]
key to display the Measurement Menu.
2
Press the
Source
softkey; then turn the Entry knob to select the first
analog channel source.
3
Press the
Type:
softkey; then, turn the Entry knob to select
Delay
.
4
Press the
Settings
softkey to select the second analog channel source
and slope for the delay measurement.
The default Delay settings measure from the rising edge of channel 1 to
the rising edge of channel 2.
Source 2
Source 1
Delay
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...