![Agilent Technologies InfiniiVision 3000 3024A MSO-X User Manual Download Page 158](http://html1.mh-extra.com/html/agilent-technologies/infiniivision-3000-3024a-mso-x/infiniivision-3000-3024a-mso-x_user-manual_2868679158.webp)
158
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
10
Triggers
Runt Trigger
The Runt trigger looks for pulses that cross one threshold but not another.
•
A positive runt pulse crosses through a lower threshold but not an
upper threshold.
•
A negative runt pulse crosses through an upper threshold but not a
lower threshold.
To trigger on runt pulses:
1
Press the
[Trigger]
key.
2
In the Trigger Menu, press the
Trigger
softkey; then, turn the Entry knob
to select
Runt
.
High level
Low level
Positive runt pulse
Negative runt pulse
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...