Math Waveforms
4
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
85
While the FFT spectrum is displayed, use the
[Math]
and
[Cursors]
keys to
switch between measurement functions and frequency domain controls in
FFT Menu.
Decreasing the effective sampling rate by selecting a greater time/div
setting will increase the low frequency resolution of the FFT display and
also increase the chance that an alias will be displayed. The resolution of
the FFT is the effective sample rate divided by the number of points in
the FFT. The actual resolution of the display will not be this fine as the
shape of the window will be the actual limiting factor in the FFTs ability
to resolve two closely space frequencies. A good way to test the ability of
the FFT to resolve two closely spaced frequencies is to examine the
sidebands of an amplitude modulated sine wave.
For the best vertical accuracy on peak measurements:
•
Make sure the probe attenuation is set correctly. The probe attenuation
is set from the Channel Menu if the operand is a channel.
•
Set the source sensitivity so that the input signal is near full screen,
but not clipped.
•
Use the Flat Top window.
•
Set the FFT sensitivity to a sensitive range, such as 2 dB/division.
For best frequency accuracy on peaks:
•
Use the Hanning window.
•
Use Cursors to place an X cursor on the frequency of interest.
•
Adjust frequency span for better cursor placement.
•
Return to the Cursors Menu to fine tune the X cursor.
For more information on the use of FFTs please refer to Agilent
Application Note 243,
The Fundamentals of Signal Analysis
at
"http://cp.literature.agilent.com/litweb/pdf/5952- 8898E.pdf"
. Additional
information can be obtained from Chapter 4 of the book
Spectrum and
Network Measurements
by Robert A. Witte.
N O T E
FFT Resolution
The FFT resolution is the quotient of the sampling rate and the number of FFT points (f
S
/N).
With a fixed number of FFT points (up to 65,536), the lower the sampling rate, the better the
resolution.
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...