Display Settings
8
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
131
To set or clear persistence
With persistence, the oscilloscope updates the display with new
acquisitions, but does not immediately erase the results of previous
acquisitions. All previous acquisitions are displayed with reduced
intensity. New acquisitions are shown in their normal color with normal
intensity.
Waveform persistence is kept only for the current display area; you cannot
pan and zoom the persistence display.
To use persistence:
1
Press the
[Display]
key.
2
Press
Persistence
; then, turn the Entry knob to select between:
•
Off
— turns off persistence.
When persistence is off, you can press the
Capture Waveforms
softkey
to perform a single- shot infinite persistence. A single acquisition's
data is displayed with reduced intensity, and it remains on the
display until you clear persistence or clear the display.
•
∞
Persistence — (infinite presistence) Results of previous acquisitions
are never erased.
Use infinite persistence to measure noise and jitter, to see the
worst- case extremes of varying waveforms, to look for timing
violations, or to capture events that occur infrequently.
•
Variable Persistence
— Results of previous acquisitions are erased after
a certain amount of time.
Variable persistence gives you a view of acquired data that is similar
to analog oscilloscopes.
When variable persistence is selected, press the
Time
softkey and use
the Entry knob to specify the amount of time that previous
acquisitions are to be displayed.
The display will begin accumulating multiple acquisitions.
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...