Vertical Controls
3
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
65
When AC coupling is selected (see
"To specify channel coupling"
on
page 63) or excessive voltage is applied to the input, the oscilloscope
automatically switches to
1M Ohm
mode to prevent possible damage.
•
1M Ohm
— is for use with many passive probes and for
general- purpose measurements. The higher impedance minimizes the
loading effect of the oscilloscope on the device under test.
This impedance matching gives you the most accurate measurements
because reflections are minimized along the signal path.
See Also
•
For more information on probing, visit:
"www.agilent.com/find/scope_probes"
•
Information about selecting a probe can be found in document number
"
Agilent Oscilloscope Probes and Accessories Selection Guide
(part
number 5989- 6162EN)", available at
"www.agilent.com"
.
To specify bandwidth limiting
1
Press the desired channel key.
2
In the Channel Menu, press the
BW Limit
softkey to enable or disable
bandwidth limiting.
When bandwidth limit is on, the maximum bandwidth for the channel is
approximately 20 MHz. For waveforms with frequencies below this, turning
bandwidth limit on removes unwanted high frequency noise from the
waveform. The bandwidth limit also limits the trigger signal path of any
channel that has
BW Limit
turned on.
To change the vertical scale knob's coarse/fine adjustment setting
1
Push the channel's vertical scale knob (or press the channel key and
then the
Fine
softkey in the Channel Menu) to toggle between fine and
coarse adjustment of the vertical scale.
When
Fine
adjustment is selected, you can change the channel's vertical
sensitivity in smaller increments. The channel sensitivity remains fully
calibrated when
Fine
is on.
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...