150
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
10
Triggers
Pulse width
trigger < qualifier
time set softkey
•
When the less than (
<
) qualifier is selected, the Entry knob sets the
oscilloscope to trigger on a pulse width less than the time value
displayed on the softkey.
•
When the time range (
><
) qualifier is selected, the Entry knob sets the
upper time range value.
Pulse width
trigger > qualifier
time set softkey
•
When the greater than (
>
) qualifier is selected, the Entry knob sets the
oscilloscope to trigger on a pulse width greater than the time value
displayed on the softkey.
•
When the time range (
><
) qualifier is selected, the Entry knob sets the
lower time range value.
Pattern Trigger
The Pattern trigger identifies a trigger condition by looking for a specified
pattern. This pattern is a logical AND combination of the channels. Each
channel can have a value of 0 (low), 1 (high), or don't care (X). A rising
or falling edge can be specified for one channel included in the pattern.
You can also trigger on a hex bus value as described on
"Hex Bus Pattern
Trigger"
on page 153.
1
Press the
[Trigger]
key.
2
In the Trigger Menu, press the
Trigger
softkey; then, turn the Entry knob
to select
Pattern
.
3
Press the
Qualifier
softkey; then, turn the Entry knob to select from the
pattern duration qualifier options:
•
Entered
– when the pattern is entered.
•
<
(Less Than) – when the pattern is present for less than a time
value.
•
>
(Greater Than) – when the pattern is present for greater than a
time value. The trigger occurs when the pattern exits (not when the
> softkey time value is exceeded).
•
Timeout
– when the pattern is present for greater than a time value.
In this case, the trigger occurs when the > softkey time value is
exceeded (not when the pattern exits).
•
>< (In Range)
– when the pattern is present for a time within a range
of values.
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...