Measurements
14
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
237
The following statistics are shown: Name of the measurement, current
measured value, mean, minimum measured value, maximum measured
value, standard deviation, and the number of times the measurement has
been made (count). Statistics are based on the total number of captured
waveforms (count).
The standard deviation shown in Statistics is calculated using the same
formula used for calculating the standard deviation measurement. The
formula is shown in the section titled
"AC RMS"
on page 223.
The source channel of the measurement is shown in parenthesis after the
measurement name. For example: "
Freq(1)
" indicates a frequency
measurement on channel 1.
You can switch the statistics
Display On
or
Display Off
. Statistics continue to
accumulate even when the statistics display is off.
When you navigate away from the Measurements Menu, the statistics
display will no longer be displayed, but statistical data collection will
continue. Return to the Measurements Menu to see the data again.
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...