Trigger Mode/Coupling
11
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
183
The external trigger input impedance is 1M Ohm. This lets you use passive
probes for general- purpose measurements. The higher impedance
minimizes the loading effect of the oscilloscope on the device under test.
To set the EXT TRIG IN units and probe attenuation:
1
Press the
[Mode/Coupling]
key in the Trigger section of the front panel.
2
In the Trigger Mode and Coupling Menu, press the
External
softkey.
3
In the External Trigger Menu, press the
Units
softkey to select between:
•
Volts
— for a voltage probe.
•
Amps
— for a current probe.
Measurement results, channel sensitivity, and the trigger level will
reflect the measurement units you have selected.
4
Press the
Probe
softkey; then, turn the entry knob to specify the probe
attenuation.
The attenuation factor can be set from 0.1:1 to 1000:1 in a 1- 2- 5
sequence.
The probe attenuation factor must be set properly for measurements to
be made correctly.
C A U T I O N
Maximum voltage at oscilloscope external trigger input
CAT I 300 Vrms, 400 Vpk; transient overvoltage 1.6 kVpk
1 M ohm input: For steady-state sinusoidal waveforms derate at 20 dB/decade above
57 kHz to a minimum of 5 Vpk
With N2863A 10:1 probe: CAT I 600 V, CAT II 300 V (DC + peak AC)
With 10073C or 10074C 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...