132
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
8
Display Settings
3
To erase the results of previous acquisitions from the display, press the
Clear Persistence
softkey.
The oscilloscope will start to accumulate acquisitions again.
4
To return the oscilloscope to the normal display mode, turn off
persistence; then, press the
Clear Persistence
softkey.
Turning off persistence does not clear the display. The display is
cleared if you press the
Clear Display
softkey or if you press the
[AutoScale]
key (which also turns off persistence).
For another method of seeing worst- case extremes of varying waveforms,
see
"Glitch or Narrow Pulse Capture"
on page 193.
To clear the display
1
Press
[Display] > Clear Display
.
You can also configure the
[Quick Action]
key to clear the display. See
"Configuring the [Quick Action] Key"
on page 298.
To select the grid type
When the
Video
trigger type is selected (see
"Video Trigger"
on page 161),
and the vertical scaling of at least one displayed channel is 140 mV/div,
the
Grid
softkey lets you select from these grid types:
•
Full
— the normal oscilloscope grid.
•
mV
— shows vertical grids, labeled on the left, from - 0.3 V to 0.8 V.
•
IRE
— (Institute of Radio Engineers) shows vertical grids in IRE units,
labeled on the left, from - 40 to 100 IRE. The 0.35 V and 0.7 V levels
from the
mV
grid are also shown and labeled at the right. When the
IRE
grid is selected, cursor values are also shown in IRE units. (Cursor
values via the remote interface are not in IRE units.)
The
mV
and
IRE
grid values are accurate (and match Y cursor values)
when the vertical scaling is 140 mV/division and the vertical offset is
245 mV.
To select the grid type:
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...