I2C/SPI Triggering and Serial Decode
24
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
373
Clock Timeout
— Select
Clock Timeout
in the
Frame by
softkey, then
select the
Timeout
softkey and turn the Entry knob to set the
minimum time that the Clock signal must be idle (not
transitioning) before the oscilloscope will search for the Data
pattern on which to trigger.
The Timeout value can be set anywhere from 100 ns to 10 s.
When you press the
Frame by
softkey, the
Display Info
graphic changes
to show timeout selection or the current state of the chip select
signal.
b
Press the
Threshold
softkey; then, turn the Entry knob to select the
chip select signal threshold voltage level.
The threshold voltage level is used in decoding, and it will become
the trigger level when the trigger type is set to the selected serial
decode slot.
When
Display Info
is enabled, information about the selected signal sources
and their threshold voltage levels, as well as a waveform diagram, appears
on the screen.
SPI Triggering
To set up the oscilloscope to capture SPI signals, see
"Setup for SPI
Signals"
on page 370.
After the oscilloscope has been set up to capture SPI signals, you can then
trigger on a data pattern that occurs at the start of a frame. The serial
data string can be specified to be from 4 to 32 bits long.
When you select the SPI trigger type and
Display Info
is enabled, a graphic
is displayed showing the current state of the frame signal, clock slope,
number of data bits, and data bit values.
1
Press
[Trigger]
.
2
In the Trigger Menu, press the
Trigger
softkey; then, turn the Entry knob
to select the serial slot (Serial 1 or Serial 2) on which the SPI signals
are being decoded.
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...