Acquisition Control
Oscilloscope Bandwidth and Sample Rate
An oscilloscope's bandwidth is typically described as the lowest frequency
at which input signal sine waves are attenuated by 3 dB (- 30% amplitude
error).
At the oscilloscope bandwidth, sampling theory says the required sample
rate is f
S
= 2f
BW
. However, the theory assumes there are no frequency
components above f
MAX
(f
BW
in this case) and it requires a system with
an ideal brick- wall frequency response.
However, digital signals have frequency components above the fundamental
frequency (square waves are made up of sine waves at the fundamental
frequency and an infinite number of odd harmonics), and typically, for
1 Ghz bandwidths and below, oscilloscopes have a Gaussian frequency
response.
Figure 34
Theoretical Brick-Wall Frequency Response
f
S
f
N
-3dB
Attenuation
Frequency
0dB
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...