Acquisition Control
12
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
189
So, in practice, an oscilloscope's sample rate should be four or more times
its bandwidth: f
S
= 4f
BW
. This way, there is less aliasing, and aliased
frequency components have a greater amount of attenuation.
See Also
Evaluating Oscilloscope Sample Rates vs. Sampling Fidelity: How to
Make the Most Accurate Digital Measurements
, Agilent Application Note
1587 (
"http://cp.literature.agilent.com/litweb/pdf/5989- 5732EN.pdf"
)
Oscilloscope Rise Time
Closely related to an oscilloscope's bandwidth specification is its rise time
specification. Oscilloscopes with a Gaussian- type frequency response have
an approximate rise time of 0.35/f
BW
based on a 10% to 90% criterion.
An oscilloscope's rise time is not the fastest edge speed that the
oscilloscope can accurately measure. It is the fastest edge speed the
oscilloscope can possibly produce.
Figure 35
Sample Rate and Oscilloscope Bandwidth
Limiting oscilloscope bandwidth (f
BW
) to 1/4 the sample rate (f
S
/4)
reduces frequency components above the Nyquist frequency (f
N
).
f
S
f
N
f
S
/4
-3dB
Attenuation
Aliased frequency
components
Frequency
0dB
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...