Aliased data is highlighted in pink. When this happens, decrease the
horizontal time/div setting and run again.
Searching for MIL-STD-1553 Data in the Lister
The oscilloscope's search capability lets you search for (and mark) certain
types of MIL- STD- 1553 data in the Lister. You can use the
[Navigate]
key
and controls to navigate through the marked rows.
1
With MIL- STD- 1553 selected as the serial decode mode, press
[Search]
.
2
In the Search Menu, press the
Search
softkey; then, turn the Entry knob
to select the serial slot (Serial 1 or Serial 2) on which the
MIL- STD- 1553 signal is being decoded.
3
Press
Search
; then, select from these options:
•
Data Word Start
— finds the start of a Data word (at the end of a valid
Data Sync pulse).
•
Command/Status Word Start
— finds the start of Command/Status word
(at the end of a valid C/S Sync pulse).
•
Remote Terminal Address
— finds the Command/Status word whose RTA
matches the specified value. The value is specified in hex.
When this option is selected, the
RTA
softkey becomes available and
lets you select the hex Remote Terminal Address value to find.
•
Remote Terminal A 11 Bits
— finds the RTA and the remaining 11
bits that match the specified criteria.
When this option is selected, these softkeys become available:
•
The
RTA
softkey lets you select the hex Remote Terminal Address
value.
•
The
Bit Time
softkey lets you select the bit time position.
•
The
0 1 X
softkey lets you set the bit time position value as a 1, 0,
or X (don't care).
•
Parity Error
— finds (odd) parity bits that are incorrect for the data in
the word.
•
Sync Error
— finds invalid Sync pulses.
•
Manchester Error
— finds Manchester encoding errors.
For more information on searching data, see
"Searching Lister Data"
on
page 126.
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...