For passive probes (such as the N2862A/B, N2863A/B, N2889A, N2890A,
10073C, 10074C, or 1165A probes), the
Probe Check
softkey appears; it
guides you through the process of compensating probes.
For some active probes (such as InfiniiMax probes), the oscilloscope can
accurately calibrate its analog channels for the probe. When you
connect a probe that can be calibrated, the
Calibrate Probe
softkey
appears (and the probe attenuation softkey may change). See
"To
calibrate a probe"
on page 68.
See Also
•
"To specify the channel units"
on page 67
•
"To specify the probe attenuation"
on page 67
•
"To specify the probe skew"
on page 68
To specify the channel units
1
Press the probe's associated channel key.
2
In the Channel Menu, press
Probe
.
3
In the Channel Probe Menu, press
Units
; then, select:
•
Volts
— for a voltage probe.
•
Amps
— for a current probe.
Channel sensitivity, trigger level, measurement results, and math functions
will reflect the measurement units you have selected.
To specify the probe attenuation
This is set automatically if the oscilloscope can identify the connected
probe. See Analog channel inputs (see
page 39
).
The probe attenuation factor must be set properly for accurate
measurement results.
If you connect a probe that is not automatically identified by the
oscilloscope, you can manually set the attenuation factor as follows:
1
Press the channel key.
2
Press the
Probe
softkey until you have selected how you want to specify
the attenuation factor, choosing either
Ratio
or
Decibels
.
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...