MIL-STD-1553/ARINC 429 Triggering and Serial Decode
26
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
393
MIL-STD-1553 Triggering
To set up the oscilloscope to capture a MIL- STD- 1553 signal, see
"Setup
for MIL- STD- 1553 Signals"
on page 391.
To set up a MIL- STD- 1553 trigger:
1
Press
[Trigger]
.
2
In the Trigger Menu, press the
Trigger
softkey; then, turn the Entry knob
to select the serial slot (Serial 1 or Serial 2) on which the
MIL- STD- 1553 signal is being decoded.
3
Press the
Trigger
softkey; then, turn the Entry knob to select the trigger
condition:
•
Data Word Start
— triggers on the start of a Data word (at the end of a
valid Data Sync pulse).
•
Data Word Stop
— triggers on the end of a Data word.
•
Command/Status Word Start
— triggers on the start of Command/Status
word (at the end of a valid C/S Sync pulse).
•
Command/Status Word Stop
— triggers on the end of a Command/Status
word.
•
Remote Terminal Address
— triggers if the RTA of the Command/Status
word matches the specified value.
When this option is selected, the
RTA
softkey becomes available and
lets you select the hex Remote Terminal Address value to trigger on.
If you select 0xXX (don't cares), the oscilloscope will trigger on any
RTA.
•
Remote Terminal A 11 Bits
— triggers if the RTA and the
remaining 11 bits match the specified criteria.
When this option is selected, these softkeys become available:
•
The
RTA
softkey lets you select the hex Remote Terminal Address
value.
•
The
Bit Time
softkey lets you select the bit time position.
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...