Digital Channels
6
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
119
Increasing the ground inductance (L), increasing the current (di) or
decreasing the transition time (dt), will all result in increasing the voltage
(V). When this voltage exceeds the threshold voltage defined in the
oscilloscope, a false data measurement will occur.
Sharing one probe ground with many probes forces all the current that
flows into each probe to return through the same common ground
inductance of the probe whose ground return is used. The result is
increased current (di) in the above equation, and, depending on the
transition time (dt), the common mode voltage may increase to a level that
causes false data generation.
In addition to the common mode voltage,longer ground returns also
degrade the pulse fidelity of the probe system. Rise time is increased, and
ringing, due to the undamped LC circuit at the input of the probe, is also
increased. Because the digital channels display reconstructed waveforms,
they do not show ringing and perturbations. You will not find ground
problems through examination of the waveform display. In fact, it is likely
Figure 24
Common Mode Input Voltage Model
Z
in
Z
in
i
n
i
2
+i
2
i
1
i
1
+i
n
+i
n
Z
in
Probe 1
Probe
Ground
Probe 2
Probe N
L (GND)
Vn (Common Mode)
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...