![Agilent Technologies InfiniiVision 3000 3024A MSO-X User Manual Download Page 294](http://html1.mh-extra.com/html/agilent-technologies/infiniivision-3000-3024a-mso-x/infiniivision-3000-3024a-mso-x_user-manual_2868679294.webp)
294
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
19
Utility Settings
To perform user calibration
Perform user- calibration:
•
Every two years or after 4000 hours of operation.
•
If the ambient temperature is >10° C from the calibration temperature.
•
If you want to maximize the measurement accuracy.
The amount of use, environmental conditions, and experience with other
instruments help determine if you need shorter User Cal intervals.
User Cal performs an internal self- alignment routine to optimize the signal
path in the oscilloscope. The routine uses internally generated signals to
optimize circuits that affect channel sensitivity, offset, and trigger
parameters.
Performing User Cal will invalidate your Certificate of Calibration. If NIST
(National Institute of Standards and Technology) traceability is required,
perform the "Performance Verification" procedure in the
Agilent
InfiniiVision 2000/3000 X- Series Oscilloscopes Service Guide
using
traceable sources.
To perform user calibration:
1
Disconnect all inputs from the front and rear panels, including the
digital channels cable on an MSO, and allow the oscilloscope to warm
up before performing this procedure.
2
Press the rear- panel CAL button to disable calibration protection.
3
Connect short (12 inch maximum) equal length cables to each analog
channel's BNC connector on the front of the oscilloscope. You will need
two equal- length cables for a 2- channel oscilloscope or four
equal- length cables for a 4- channel oscilloscope.
Use 50W RG58AU or equivalent BNC cables when performing User Cal.
For a 2- channel oscilloscope, connect a BNC tee to the equal length
cables. Then connect a BNC(f)- to- BNC(f) (also called a barrel
connector) to the tee as shown below.
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...