238
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
14
Measurements
To reset the statistics measurements, press the
Reset Statistics
softkey. This
resets all statistics and begins recording statistical data again.
Each time a new measurement (for example: frequency, period, or
amplitude) is added the statistics are reset and accumulation of statistical
data begins again.
When the
[Single]
key is pressed, statistics are reset and a single
measurement is done (count = 1). Successive
[Single]
acquisitions
accumulate statistical data (and the count is incremented).
Press the
Transparent
softkey to disable the Transparent mode. This shows
statistics with a gray background. Press the
Transparent
softkey again to
enable Transparent mode. This writes measurement values, statistics, and
cursor values on the screen without a background. The Transparent
setting affects measurement statistics, reference waveform information,
and the optional Mask test feature's statistics display.
Relative
σ
— When enabled, the standard deviation shown in measurement
statistics becomes a relative standard deviation, that is, standard
deviation/mean.
Max Count
— This softkey specifies the number of values used when
calculating measurement statistics.
The
Increment Statistics
softkey only appears when the acquisition is stopped
and the optional segmented memory feature is off. Press the
[Single]
or
[Run/Stop]
key to stop the acquisition. You can use the horizontal position
control (in the Horizontal control section of the front panel) to pan
through the waveform. Active measurements will stay on screen, allowing
you to measure various aspects of the captured waveforms. Press
Increment
Statistics
to add the currently measured waveform to the collected
statistical data.
The
Analyze Segments
softkey only appears when the acquisition is stopped
and the optional segmented memory feature is on. After an acquisition
has completed (and the oscilloscope is stopped), you can press the
Analyze
Segments
softkey to accumulate measurement statistics for the acquired
segments.
You can also turn on infinite persistence (in the Display Menu) and press
the
Analyze Segments
softkey to create an infinite persistence display.
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...