![Agilent Technologies InfiniiVision 3000 3024A MSO-X User Manual Download Page 227](http://html1.mh-extra.com/html/agilent-technologies/infiniivision-3000-3024a-mso-x/infiniivision-3000-3024a-mso-x_user-manual_2868679227.webp)
Measurements
14
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
227
Counter
The InfiniiVision 3000 X- Series oscilloscopes have an integrated hardware
frequency counter which counts the number of cycles that occur within a
period of time (known as the gate time) to measure the frequency of a
signal.
The gate time for the Counter measurement is automatically adjusted to
be 100 ms or twice the current time window, whichever is longer, up to 1
second.
The Counter can measure frequencies up to the bandwidth of the
oscilloscope. The minimum frequency supported is 1/(2 X gate time).
The hardware counter uses the trigger comparator output. Therefore, the
counted channel's trigger level (or threshold for digital channels) must be
set correctly. The Y cursor shows the threshold level used in the
measurement.
Analog and digital channels can be selected as the source.
Only one Counter measurement can be displayed at a time.
Figure 47
Isolating an event for Frequency measurement
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...