372
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
24
I2C/SPI Triggering and Serial Decode
9
Press the
CS
softkey to open the SPI Chip Select Menu.
In the SPI Chip Select Menu:
a
Press the
Frame by
softkey to select a framing signal that the
oscilloscope will use for determining which clock edge is the first
clock edge in the serial stream.
You can set the oscilloscope to trigger during a high chip select (
CS
),
a low chip select (
~CS
), or after a
Timeout
period during which the
clock signal has been idle.
•
If the framing signal is set to
CS
(or
~CS
), the first clock edge as
defined, rising or falling, seen after the
CS
(or
~CS
) signal
transitions from low to high (or high to low) is the first clock in
the serial stream.
Chip Select
— Press the
CS
or
~CS
softkey; then, turn the Entry
knob to select the channel that is connected to the SPI frame line.
The label (~CS or CS) for the source channel is automatically set.
The data pattern and the clock transition must occur during the
time when the framing signal is valid. The framing signal must be
valid for the entire data pattern.
•
If the framing signal is set to
Timeout
, the oscilloscope generates
it's own internal framing signal after it sees inactivity on the serial
clock line.
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...