Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
10
Triggers
The
Generic
selection lets you trigger on custom bi- level and tri- level
sync video standards. See
"To set up Generic video triggers"
on
page 166.
7
Press the
Auto Setup
softkey to automatically set up the oscilloscope for
the selected
Source
and
Standard
:
•
Source channel vertical scaling is set to 140 mV/div.
•
Source channel offset is set to 245 mV.
•
Source channel is turned on.
•
Trigger type is set to
Video
.
•
Video trigger mode is set to
All Lines
(but left unchanged if
Standard
is
Generic
).
•
Display
Grid
type is set to
IRE
(when
Standard
is
NTSC
) or
mV
(see
"To
select the grid type"
on page 132).
•
Horizontal time/division is set to 10 µs/div for NTSC/PAL/SECAM
standards or 4 µs/div for EDTV or HDTV standards (unchanged for
Generic
).
•
Horizontal delay is set so that trigger is at first horizontal division
from the left (unchanged for
Generic
).
You can also press
[Analyze]> Features
and then select
Video
to quickly
access the video triggering automatic set up and display options.
8
Press the
Mode
softkey to select the portion of the video signal that you
would like to trigger on.
The Video trigger modes available are:
•
Field1
and
Field2
— Trigger on the rising edge of the first serration
pulse of field 1 or field 2 (interlaced standards only).
•
All Fields
— Trigger on the rising edge of the first pulse in the vertical
sync interval.
•
All Lines
— Trigger on all horizontal sync pulses.
•
Line
— Trigger on the selected line number (EDTV and HDTV
standards only).
•
Line: Field1
and
Line:Field2
— Trigger on the selected line # in field 1 or
field 2 (interlaced standards only).
•
Line: Alternate
— Alternately trigger on the selected line # in field 1
and field 2 (NTSC, PAL, PAL- M, and SECAM only).
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...