242
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
15
Mask Testing
Run Until
The Run Until softkey lets you specify a condition on which to terminate
testing.
• Forever — The oscilloscope runs continuously. However, if an error occurs
the action specified using the On Error softkey will occur.
• Minimum # of Tests — Choose this option and then use the # of Tests
softkey to select the number of times the oscilloscope will trigger, display
the waveform(s), and compare them to the mask. The oscilloscope will
stop after the specified number of tests have been completed. The
specified minimum number of tests may be exceeded. If an error occurs the
action specified using the On Error softkey will occur. The actual number of
tests completed is displayed above the softkeys.
• Minimum Time — Choose this option and then use the Test Time softkey
to select how long the oscilloscope will run. When the selected time has
passed the oscilloscope will stop. The specified time may be exceeded. If
an error occurs the action specified using the On Error softkey will occur.
The actual test time is displayed above the softkeys.
• Minimum Sigma — Choose this option and then use the Sigma softkey to
select a minimum sigma. The mask test runs until enough waveforms are
tested to achieve a minimum test sigma. (If an error occurs the
oscilloscope will perform the action specified by the On Error softkey.)
Note that this is a test sigma (the max achievable process sigma,
assuming no defects, for a certain number of tested waveforms) as
opposed to a process sigma (which is tied to the amount of failures per
test). The sigma value may exceed the selected value when a small sigma
value is chosen. The actual sigma is displayed.
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...