364
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
24
I2C/SPI Triggering and Serial Decode
•
Frame (Start: Addr7: Read: Ack: Data)
or
Frame (Start: Addr7: Write: Ack: Data)
—
The oscilloscope triggers on a read or write frame in 7- bit addressing
mode on the 17th clock edge if all bits in the pattern match. For
triggering purposes, a restart is treated as a start condition.
•
Frame (Start: Addr7: Read: Ack: Data: Ack: Data2)
or
Frame (Start: Addr7: Write:
Ack: Data: Ack: Data2)
— The oscilloscope triggers on a read or write
frame in 7- bit addressing mode on the 26th clock edge if all bits in
the pattern match. For triggering purposes, a restart is treated as a
start condition.
•
10-bit Write
— The oscilloscope triggers on a 10- bit write frame on the
26th clock edge if all bits in the pattern match. The frame is in the
format:
Frame (Start: Address byte 1: Write: Address byte 2: Ack: Data)
For triggering purposes, a restart is treated as a start condition.
SDA
SCL
Control
byte
Read
R/
Ack
Data
Ack
Trigger point
Start or
Restart
Condition
SDA
SCL
Stop
Condition
Start or
Restart
Condition
Trigger point
17th clock edge
Address
Read
Write
R/
Ack
Data
Ack
Read
Write
Start
Condition
Address
R/W
Ack
Data
Ack
Data 2
Trigger point
26th clock edge
Ack
Stop
Condition
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...