Acquisition Control
12
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
195
Use the horizontal position knob (
) to pan through the waveform to set
the expanded portion of the normal window around the glitch.
Averaging Acquisition Mode
The Averaging mode lets you average multiple acquisitions together to
reduce noise and increase vertical resolution (at all time/div settings).
Averaging requires a stable trigger.
The number of averages can be set from 2 to 65536 in power- of- 2
increments.
A higher number of averages reduces noise more and increases vertical
resolution.
The higher the number of averages, the slower the displayed waveform
responds to waveform changes. You must compromise between how quickly
the waveform responds to changes and how much you want to reduce the
displayed noise on the signal.
To use the Averaging mode:
1
Press the
[Acquire]
key, then press the
Acq Mode
softkey until the
Averaging mode is selected.
2
Press the
#Avgs
softkey and turn the Entry knob to set the number of
averages that best eliminates the noise from the displayed waveform.
The number of acquisitions being averaged is displayed in the
# Avgs
softkey.
# Avgs
Bits of resolution
2
8
4
9
16
10
64
11
≥
256
12
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...