starts on Line 4. In the case of NTSC video, the oscilloscope will trigger
on color field 1 alternating with color field 3 (see the following figure).
This setup can be used to measure the envelope of the reference burst.
If a more detailed analysis is required, then only one color field should be
selected to be the trigger. You can do this by using the
Field Holdoff
softkey
in the Video Trigger Menu. Press the
Field Holdoff
softkey and use the Entry
knob to adjust the holdoff in half- field increments until the oscilloscope
triggers on only one phase of the color burst.
A quick way to synchronize to the other phase is to briefly disconnect the
signal and then reconnect it. Repeat until the correct phase is displayed.
When holdoff is adjusted using the
Field Holdoff
softkey and the Entry knob,
the corresponding holdoff time will be displayed in the Trigger Mode and
Coupling Menu.
Figure 31
Triggering on Color Field 1 Alternating with Color Field 3
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...