I2C/SPI Triggering and Serial Decode
24
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
365
5
If you have set the oscilloscope to trigger on an EEPROM Data Read
condition:
Press the
Data is
softkey to set the oscilloscope to trigger when data is
= (equal to),
≠
(not equal to), < (less than), or > (greater than) the data
value set in the
Data
softkey.
The oscilloscope will trigger on the clock edge for the Ack bit after the
trigger event is found. This data byte does not need to occur directly
after the control byte. The oscilloscope will trigger on any data byte
that meets the criteria defined by the
Data is
and
Data
softkeys during a
current address read or a random read or a sequential read cycle.
6
If you have set the oscilloscope to trigger on a 7- bit address read or
write frame condition or a 10- bit write frame condition:
a
Press the
Address
softkey and turn the Entry knob to select the 7- bit
or 10- bit device address.
You can select from an address range of 0x00 to 0x7F (7- bit) or
0x3FF (10- bit) hexadecimal. When triggering on a read/write frame,
the oscilloscope will trigger after the start, address, read/write,
acknowledge, and data events occur.
If don't care is selected (0xXX or 0xXXX) for the address, the
address will be ignored. The trigger will always occur on the 17th
clock for 7- bit addressing or 26th clock for 10- bit addressing.
b
Press the
Data
value softkey and turn the Entry knob to select the
8- bit data pattern on which to trigger.
You can select a data value in the range of 0x00 to 0xFF
(hexadecimal). The oscilloscope will trigger after the start, address,
read/write, acknowledge, and data events occur.
SCL
SDA
Stop
Condition
Start or
Restart
Condition
Trigger point
26th clock edge
Address
1st byte
R/
Ack1
Ack
Address
2nd byte
Write
Ack2
Data
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...