![Agilent Technologies InfiniiVision 3000 3024A MSO-X User Manual Download Page 352](http://html1.mh-extra.com/html/agilent-technologies/infiniivision-3000-3024a-mso-x/infiniivision-3000-3024a-mso-x_user-manual_2868679352.webp)
352
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
23
FlexRay Triggering and Serial Decode
6
Press
Source
and select the analog channel that is probing the FlexRay
signal.
7
Press
Threshold
; then, turn the Entry knob to set the threshold voltage
level.
The threshold level should be set below the idle level.
The threshold voltage level is used in decoding and will become the
trigger level when the trigger type is set to the selected serial decode
bus.
8
Press
Baud
and select the baud rate of the FlexRay signal being probed.
9
Press
Bus
and select the bus type of the FlexRay signal being probed.
It is important to specify the correct bus because this setting affects
CRC error detection.
10
Press
Auto Setup
to perform the following actions:
•
Sets the selected source channel's impedance to 50 Ohms, assuming a
differential active probe that requires a 50 ohm termination is being
used .
•
Sets the selected source channel's probe attenuation to 10:1.
•
Sets the trigger level (on the selected source channel) to - 300 mV.
•
Turns on trigger Noise Reject.
•
Turns on Serial Decode.
•
Sets the trigger type to FlexRay.
FlexRay Triggering
To set up the oscilloscope to capture a FlexRay signal, see
"Setup for
FlexRay Signals"
on page 351.
After you have set up the oscilloscope to capture a FlexRay signal, you
can then set up triggers on frames (see
page 353
), errors (see
page 354
),
or events (see
page 354
).
N O T E
To display FlexRay serial decode, see
"FlexRay Serial Decode"
on page 355.
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...