![Agilent Technologies InfiniiVision 3000 3024A MSO-X User Manual Download Page 68](http://html1.mh-extra.com/html/agilent-technologies/infiniivision-3000-3024a-mso-x/infiniivision-3000-3024a-mso-x_user-manual_2868679068.webp)
68
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
3
Vertical Controls
3
Turn the Entry knob
to set the attenuation factor for the
connected probe.
When measuring voltage values, the attenuation factor can be set from
0.1:1 to 1000:1 in a 1- 2- 5 sequence.
When measuring current values with a current probe, the attenuation
factor can be set from 10 V/A to 0.001 V/A.
When specifying the attenuation factor in decibels, you can select values
from - 20 dB to 60 dB.
If Amps is chosen as the units and a manual attenuation factor is chosen,
then the units as well as the attenuation factor are displayed above the
Probe
softkey.
To specify the probe skew
When measuring time intervals in the nanoseconds (ns) range, small
differences in cable length can affect the measurement. Use
Skew
to
remove cable- delay errors between any two channels.
1
Probe the same point with both probes.
2
Press one of the probes associated channel key.
3
In the Channel Menu, press
Probe
.
4
In the Channel Probe Menu, press
Skew
; then, select the desired skew
value.
Each analog channel can be adjusted ±100 ns in 10 ps increments for a
total of 200 ns difference.
The skew setting is not affected by pressing
[Default Setup]
or
[Auto Scale]
.
To calibrate a probe
The
Calibrate Probe
softkey guides you through the process of calibrating
probes.
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...