UART/RS232 Triggering and Serial Decode
27
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
409
UART/RS232 Triggering
To set up the oscilloscope to capture UART/RS- 232 signals, see
"Setup for
UART/RS232 Signals"
on page 407.
To trigger on a UART (Universal Asynchronous Receiver/Transmitter)
signal connect the oscilloscope to the Rx and Tx lines and set up a trigger
condition. RS232 (Recommended Standard 232) is one example of a UART
protocol.
1
Press
[Trigger]
.
2
In the Trigger Menu, press the
Trigger
softkey; then, turn the Entry knob
to select the serial slot (Serial 1 or Serial 2) on which the UART/RS232
signals are being decoded.
3
Press the
Trigger Setup
softkey to open the UART/RS232 Trigger Setup
Menu.
4
Press the
Base
softkey to select Hex or ASCII as the base displayed on
the Data softkey in the UART/RS232 Trigger Setup Menu.
Note that the setting of this softkey does not affect the selected base of
the decode display.
5
Press the
Trigger
softkey and set up the desired trigger condition:
•
Rx Start Bit
— The oscilloscope triggers when a start bit occurs on Rx.
N O T E
In the serial decode display, the most significant bit is always displayed on the left
regardless of how Bit Order is set.
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...