![Agilent Technologies InfiniiVision 3000 3024A MSO-X User Manual Download Page 219](http://html1.mh-extra.com/html/agilent-technologies/infiniivision-3000-3024a-mso-x/infiniivision-3000-3024a-mso-x_user-manual_2868679219.webp)
Measurements
14
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
219
•
"Maximum"
on page 219
•
"Minimum"
on page 219
•
"Amplitude"
on page 219
•
"Top"
on page 219
•
"Base"
on page 220
•
"Overshoot"
on page 221
•
"Preshoot"
on page 222
•
"Average"
on page 222
•
"DC RMS"
on page 223
•
"AC RMS"
on page 223
•
"Ratio"
on page 225
Peak-Peak
The peak- to- peak value is the difference between Maximum and Minimum
values. The Y cursors show the values being measured.
Maximum
Maximum is the highest value in the waveform display. The Y cursor
shows the value being measured.
Minimum
Minimum is the lowest value in the waveform display. The Y cursor shows
the value being measured.
Amplitude
The Amplitude of a waveform is the difference between its Top and Base
values. The Y cursors show the values being measured.
Top
The Top of a waveform is the mode (most common value) of the upper
part of the waveform, or if the mode is not well defined, the top is the
same as Maximum. The Y cursor shows the value being measured.
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...