198
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
12
Acquisition Control
When capturing multiple infrequent trigger events it is advantageous to
divide the oscilloscope's memory into segments. This lets you capture
signal activity without capturing long periods of signal inactivity.
Each segment is complete with all analog channel, digital channel (on MSO
models), and serial decode data.
When using segmented memory, use the Analyze Segments feature (see
"Measurements, Statistics, and Infinite Persistence with Segmented
Memory"
on page 199) to show infinite persistence across all acquired
segments. See also
"To set or clear persistence"
on page 131 for details.
To acquire to
segmented
memory
1
Set up a trigger condition. (See
Chapter 10
, “Triggers,” starting on page
141 for details.)
2
Press the
[Acquire]
key in the Waveform section of the front panel.
3
Press the
Segmented
softkey.
4
In the Segmented Memory Menu, press the
Segmented
softkey to enable
segmented memory acquisitions.
5
Press the
# of Segs
softkey and turn the Entry knob to select the
number of segments into which you would like to divide the
oscilloscope's memory.
Memory can be divided into as few as two segments and as many as
1000 segments, depending on the oscilloscope model.
6
Press the
[Run]
or
[Single]
key.
The oscilloscope runs and fills a memory segment for each trigger event.
When the oscilloscope is busy acquiring multiple segments, the progress is
displayed in the upper right area of the display. The oscilloscope
continues to trigger until memory is filled, then the oscilloscope stops.
If the signal you are measuring has more than about 1 s of inactivity,
consider selecting the
Normal
trigger mode to prevent AutoTriggering. See
"To select the Auto or Normal trigger mode"
on page 178.
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...