![Agilent Technologies InfiniiVision 3000 3024A MSO-X User Manual Download Page 284](http://html1.mh-extra.com/html/agilent-technologies/infiniivision-3000-3024a-mso-x/infiniivision-3000-3024a-mso-x_user-manual_2868679284.webp)
284
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
19
Utility Settings
•
LAN
— When a DSOXLAN LAN/VGA module is installed, you can use
the
LAN Settings
and
LAN Reset
softkeys to configure the LAN
interface. See
"Setting up the Oscilloscope's LAN Connection"
on
page 284.
•
GPIB
— When a DSOXGPIB GPIB module is installed, you can use the
Address
softkey to configure the GPIB address.
•
There are no configuration settings for the USB interface.
When an I/O interface is installed, remote control over that interface is
always enabled. Also, the oscilloscope can be controlled via multiple I/O
interfaces (for example, USB and LAN) at the same time.
See Also
•
Chapter 20
, “Web Interface,” starting on page 301 (when the
oscilloscope is connected to a LAN).
•
"Remote Programming via the Web Interface"
on page 306
•
The oscilloscope's
Programmer's Guide
.
•
"Remote Programming with Agilent IO Libraries"
on page 307
Setting up the Oscilloscope's LAN Connection
When the DSOXLAN LAN/VGA module is installed, you can place the
oscilloscope on the network and set up its LAN connection. Once that is
done, you can use the oscilloscope's web interface or remotely control the
oscilloscope via the LAN interface.
The oscilloscope supports methods for automated LAN configuration or
manual LAN configuration (see
"To establish a LAN connection"
on
page 285). It is also possible to set up a point- to- point LAN connection
between a PC and the oscilloscope (see
"Stand- alone (Point- to- Point)
Connection to a PC"
on page 286).
Once the oscilloscope is set up on the network, you can use the
oscilloscope's web page to view or change its network configuration and
access additional settings (like the network password). See
Chapter 20
,
“Web Interface,” starting on page 301.
N O T E
When you connect the oscilloscope to a LAN it is a good practice to limit access to the
oscilloscope by setting a password. By default, the oscilloscope is not password protected.
See
"Setting a Password"
on page 312 to set a password.
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...