3
Advanced Troubleshooting
Replacing Components of the Ion Source
220
TSQ 9000 Hardware Manual
Thermo Scientific
.
14. Twist the handle of the source exchange tool to the right and pull it back toward you and
down the right side of the barrel’s metal track until the line at the end of the track appears.
Note
Do not push down on the handle of the source exchange tool while inserting the
ion source cartridge. This could result in opening an air leak that may cause your
instrument to shut itself down to protect the vacuum pumps.
CAUTION INSTRUMENT DAMAGE:
Make sure the vacuum interlock handle is all the way
up. Otherwise if you insert the tool with the handle down, you can damage the
instrument.
Figure 214.
Pulling the Tool Down the Right Track
Содержание TSQ 9000