Advanced Troubleshooting
Replacing Components of the Ion Source
Thermo Scientific
TSQ 9000 Hardware Manual
187
5. Twist off each source-repeller thumbscrew.
6. Remove the thumbscrew-repeller spacers from the thumbscrews.
7. Move the locking ring out of the way.
Figure 169.
Removing the Source-Repeller Thumbscrews
Figure 170.
Removing the Thumbscrew-Repeller Spacers
Source-Repeller
Thumbscrews
Thumbscrew-Repeller
Spacers