14. Replace the front manifold plate.
15. Reattach the front plate o-ring.
16. Reattach the magnet yoke, alignment collar, and source magnet retainers.
Figure 282.
Replacing the Front Manifold Plate
Figure 283.
Reattaching the Front Plate O-Ring
Figure 284.
Reattaching the Magnet Yoke and Source Magnet Retainers
Manifold
O-ring
Magnet Yoke
Source Magnet
Retainers