3
Advanced Troubleshooting
Replacing Components of the Ion Source
Thermo Scientific
TSQ 9000 Hardware Manual
217
5. Gently pull the source holder away from the ion source cartridge.
Figure 210.
Reattaching the Ion Source Cartridge to the Tool
Note
The ion volume and lenses have keying features to assure the lens spring contacts
will align with the metal parts of the lenses.
Figure 211.
Removing the Source Holder
Tip
When the source exchange tool is inverted, the source should stay attached. If this
does not happen, the source will not insert correctly into the instrument.