Advanced Troubleshooting
Replacing the Collision Gas Components
386
TSQ 9000 Hardware Manual
Thermo Scientific
Figure 410.
Disconnecting the Controller Interface Board
17. Disconnect the tubing from the tee connecting the collision gas module to the foreline
adapter. The tubing is connected at the tee with a compression fitting.
Figure 411.
Identifying the Tubes on the Collision Gas Module
18. If you need to replace the tee, unscrew it from the foreline adapter and replace it. You may
also replace the foreline adapter if necessary.
19. Disconnect the collision gas tubing from the module. The tubing is secured to the
module using a compression fitting.
Ethernet Cable
Controller
Interface Board
PC Communication
Board
Tubing to Bulkhead
Fitting on Chassis
Tubing to
Collision Cell
through Manifold
Tubing to Tee
Connected to
Foreline Adapter
Compression
Fitting