Advanced Troubleshooting
Replacing Components of the Analyzer
138
TSQ 9000 Hardware Manual
Thermo Scientific
9. Set the collision cell assembly on a clean work surface. Disconnect the Q1 entrance lens
wire from the source interface board.
Figure 105.
Removing the Analyzer Tray Screws
10. Remove the Q1 entrance lens wire from the source interface board.
11. Remove the ion guide assembly screws.
12. Remove the ion guide assembly.
13. Use a T10 Torxhead screwdriver to remove the lens heater block grounding strap from the
analyzer tray.
Figure 104.
Removing the Collision Cell and Q1 Quad Assembly from the Analyzer Tray
Collision
Cell
Assembly
Screw to
Axial Field
Wire
Q1 Entrance
Lens Endcap
Q1 Quad
Assembly
Screw to
Analyzer Tray
Q1 Entrance Lens
Endcap Screws to
Analyzer Tray
Q1 Exit Endcap
Screw
Q3 Exit Endcap
Screw
Ion Guide Screws
Q3 Entrance Endcap Screw