3
Advanced Troubleshooting
Replacing Fans and Filters
400
TSQ 9000 Hardware Manual
Thermo Scientific
Air Deflector
To replace the air deflector
1. Shut down your system by following the steps in
“Shutting Down the TSQ 9000 System”
. You do not need to remove the ion source cartridge or column or vent the
instrument. If you are using hydrogen as a carrier gas, power off the GC.
2. Slide the top cover panel toward the back of the instrument and lift it off.
3. Use a T20 Torxhead screwdriver to remove the four screws securing the left hand front
panel to the chassis. One of the screws is located on the interior front panel of the
instrument.
Figure 425.
Locating the Left Hand Front Panel Screws
4. Lift the left hand front panel off the instrument.
5. Remove the two rivets holding the air deflector to the chassis cooling fan.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for ordering information.
Screws
Left Hand Front
Panel
Screw on Front
of Instrument