3
Advanced Troubleshooting
Replacing the Collision Gas Components
Thermo Scientific
TSQ 9000 Hardware Manual
385
14. Use a T20 Torxhead screwdriver to loosen the two screws holding the support bracket to
the bottom of the instrument.
15. Remove the controller interface board and PC communication board from the
instrument. The boards are attached together.
16. Disconnect the Ethernet cable from the PC communication board.
Figure 409.
Removing the Controller Interface Board Support Bracket
Support Bracket
Controller
Interface Board