Performing Routine Maintenance
Maintenance Supplies and Tools
Thermo Scientific
TSQ 9000 Hardware Manual
3
Maintenance Supplies and Tools
To perform routine maintenance on the TSQ 9000 instrument, you will need the supplies
and tools listed in the following table. The supplies in the VPI MS Toolkit or the MS Toolkit
are included in the toolkit shipped with your instrument.
In addition to the MS Toolkit, you can also purchase a TSQ 9000 Screw Kit, which includes
extra screws, nuts, and standoffs. Refer to the
ISQ and TSQ GC-MS Spare Parts Guide l
for
ordering information.
Item
Included in
the VPI MS
Toolkit
a
a
A TSQ 9000 System with a VPI ships with the VPI MS Toolkit.
Included in
the MS
Toolkit
Not Included in
the MS Toolkit
Cloth or paper, lint-free
✓
Column measuring tool
✓
✓
Source exchange tool
✓
Small source exchange tool
✓
Forceps
✓
✓
Gas, clean and dry
✓
Gloves, clean, lint-free
✓
Protective eye wear
✓
Torx® T10 screwdriver
✓
✓
Torx® T20 screwdriver
✓
✓
Torx® T30 screwdriver
✓
✓
Wrenches, open-ended, 1/4-in.,
5/16-in., 3/8 in., 7/16-in.
✓
✓
Note
The tools in the MS Toolkit are shipped in a black plastic case with the instrument.
Each tool in the kit can be ordered separately or you can order an additional toolkit. See
the
ISQ and TSQ GC-MS Spare Parts Guide
for ordering information.
IMPORTANT
Many nitrile and latex gloves not certified for clean room use contain
silicone mold releasing agents that will contaminate the instrument. For this reason, clean
room gloves are strongly recommended. We recommend Cardinal Health CP100 Nitrile
Cleanroom Gloves. See the
ISQ and TSQ GC-MS Spare Parts Guide
for ordering
information