3
Advanced Troubleshooting
Replacing Components of the Ion Source
Thermo Scientific
TSQ 9000 Hardware Manual
195
8. Slide the source interface board and source block onto the tray without damaging or
disconnecting the wires.
9. Reattach the two screws to the back of the source block.
Figure 180.
Sliding the Ion Source Block onto the Analyzer Tray
Figure 181.
Reconnecting the Ion Source to the Analyzer Tray
IMPORTANT
When assembling the source and optics block, take extra care to ensure the
blocks are centered properly and that the ion source cartridge provided with the
instrument can easily slide in and out of the assembly.
Wires
Source Block
Source
Interface
Board
Screws