Advanced Troubleshooting
Replacing Components of the Manifold
Thermo Scientific
TSQ 9000 Hardware Manual
253
Manifold Door Hinge
To replace the manifold door hinge:
1. Shut down your system by following the steps in
“Shutting Down the TSQ 9000 System”
.
2. Slide the top cover panel toward the back of the instrument and lift it off.
3. Remove the four screws on the left hand front panel and remove the panel. One is located
at the bottom left side of the front of the instrument.
Figure 266.
Locating the Left Hand Front Panel Screws
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Note
You do not have to remove the ion source cartridge or the column.
Nuts
Screw Located on Front of
Instrument