Advanced Troubleshooting
Replacing Components of the Ion Source
Thermo Scientific
TSQ 9000 Hardware Manual
191
Figure 174.
Locating Wires on the Source Interface Board
5. Attach a new lens plate and springs to the lens heater block and reattach the screws.
6. Reattach the lens 1, lens 2, and lens 3 wires from the source interface board using a
flat-tipped screwdriver.
7. Once the components are correctly reattached, follow the instructions in
Analyzer Tray in the TSQ 9000 Instrument”
to replace the tray in the TSQ
instrument.
8. Restart your system by following the instructions in