Performing Routine Maintenance
Replacing the Electron Multiplier
Thermo Scientific
TSQ 9000 Hardware Manual
85
22. Put the electron multiplier plate back into place and reattach the four screws. Tighten
them in this order: screw 1, screw 2, screw 3, screw 4. See
for clarification.
23. Reconnect the electrometer board cable and large feedthrough cables.
Figure 76.
Reattaching the Electron Multiplier Plate
Screw 2
Screw 3
Screw 4
Screw 1
CAUTION INSTRUMENT DAMAGE:
Do not mix up the cables or you will damage the
power supply.
ATTENTION DOMMAGES À L’INSTRUMENT
: ne confondez pas les câbles ou vous
endommagerez l’alimentation.