3
Advanced Troubleshooting
Replacing the Boards in the TSQ 9000 Instrument
Thermo Scientific
TSQ 9000 Hardware Manual
339
8. Remove the controller interface board and PC communication board from the
instrument.
9. Remove the six screws holding the fan plenum to the chassis. Two are on the left side of
the instrument, and four are attached to the interior front panel.
10. Disconnect the fan from the distribution board.
11. Remove the entire fan plenum assembly including the fan and air deflector from the
instrument.
Figure 363.
Removing the Controller Interface Board Support Bracket
Support Bracket
Controller
Interface Board
PC Communication Board