1
Performing Routine Maintenance
Replacing the Electron Multiplier
84
TSQ 9000 Hardware Manual
Thermo Scientific
14. Flip the electron multiplier and plate over so that the plate is resting on the worktable.
15. Remove the stainless steel wire that from the gold connector on the electron multiplier
and move it out of the way.
16. Lift up on the spring tab at the top of the electron multiplier and remove the electron
multiplier.
17. Insert the new electron multiplier through the bottom tab first and then slide it into the
spring tab at the top until it clicks into place.
18. Make sure the electron multiplier fits snugly.
19. Reconnect the stainless steel wire to the gold connector.
20. Inspect the o-ring between the manifold and detector assembly for debris. If the o-ring is
dirty, clean or replace it.
21. Flip the electron multiplier and plate over and insert them into the instrument.
Figure 75.
Replacing the Electron Multiplier
IMPORTANT
Many nitrile and latex gloves not certified for clean room use contain
silicone mold releasing agents that will contaminate the instrument. For this reason, clean
room gloves are strongly recommended when touching components inside the vacuum
manifold. We recommend Cardinal Health CP100 Nitrile Cleanroom Gloves. See the
ISQ
and TSQ GC-MS Spare Parts Guide
for ordering information
Electron
Multiplier
Top
Spring
Tab
Stainless
Steel Wire
Gold
Connector
Bottom
Spring Tab