Advanced Troubleshooting
Replacing Components of the Vacuum Interlock
240
TSQ 9000 Hardware Manual
Thermo Scientific
4. Disconnect the vacuum interlock cable from the connector.
5. Unscrew and then lower the vacuum interlock handle.
6. Pinch together the two springs on the left side the vacuum interlock shield and remove it.
Figure 245.
Disconnecting the Vacuum Interlock Cable
Figure 246.
Removing the Vacuum Interlock Shield
Vacuum
Interlock
Cable
Connector
Springs
Vacuum
Interlock
Handle
Lower the Vacuum Interlock Handle